![]() |
![]() |
||||
FDA GMP and GLP regulations
21CFR PART
820 QUALITY SYSTEM REGULATION
Sec.
820.1 Scope. (a)Applicability.
(1) Current good manufacturing practice (CGMP) requirements are set forth in
this quality system regulation. The requirements in this part govern the
methods used in, and the facilities and controls used for, the design,
manufacture, packaging, labeling, storage, installation, and servicing of all
finished devices intended for human use. The requirements in this part are
intended to ensure that finished devices will be safe and effective and
otherwise in compliance with the Federal Food, Drug, and Cosmetic Act (the
act). This part establishes basic requirements applicable to manufacturers of
finished medical devices. If a manufacturer engages in only some operations
subject to the requirements in this part, and not in others, that manufacturer
need only comply with those requirements applicable to the operations in which
it is engaged. With respect to class I devices, design controls apply only to
those devices listed in 820.30(a)(2). This regulation does not apply to
manufacturers of components or parts of finished devices, but such
manufacturers are encouraged to use appropriate provisions of this regulation
as guidance. Manufacturers of human blood and blood components are not subject
to this part, but are subject to part 606 of this chapter. Manufacturers of
human cells, tissues, and cellular and tissue-based products (HCT/Ps), as
defined in 1271.3(d) of this chapter, that are medical devices (subject to
premarket review or notification, or exempt from notification, under an
application submitted under the device provisions of the act or under a
biological product license application under section 351 of the Public Health
Service Act) are subject to this part and are also subject to the donor-eligibility
procedures set forth in part 1271 subpart C of this chapter and applicable
current good tissue practice procedures in part 1271 subpart D of this chapter.
In the event of a conflict between applicable regulations in part 1271 and in
other parts of this chapter, the regulation specifically applicable to the
device in question shall supersede the more general. (2) The
provisions of this part shall be applicable to any finished device as defined
in this part, intended for human use, that is manufactured, imported, or
offered for import in any State or Territory of the United States, the District
of Columbia, or the Commonwealth of Puerto Rico. (3) In
this regulation the term "where appropriate" is used several times.
When a requirement is qualified by "where appropriate," it is deemed
to be "appropriate" unless the manufacturer can document
justification otherwise. A requirement is "appropriate" if
nonimplementation could reasonably be expected to result in the product not
meeting its specified requirements or the manufacturer not being able to carry
out any necessary corrective action. (b) The
quality system regulation in this part supplements regulations in other parts
of this chapter except where explicitly stated otherwise. In the event of a conflict
between applicable regulations in this part and in other parts of this chapter,
the regulations specifically applicable to the device in question shall
supersede any other generally applicable requirements. (c)Authority.
Part 820 is established and issued under authority of sections 501, 502, 510,
513, 514, 515, 518, 519, 520, 522, 701, 704, 801, 803 of the act (21 U.S.C.
351, 352, 360, 360c, 360d, 360e, 360h, 360i, 360j, 360l, 371, 374, 381, 383).
The failure to comply with any applicable provision in this part renders a
device adulterated under section 501(h) of the act. Such a device, as well as
any person responsible for the failure to comply, is subject to regulatory
action. (d)Foreign
manufacturers. If a manufacturer who offers devices for import into the United
States refuses to permit or allow the completion of a Food and Drug
Administration (FDA) inspection of the foreign facility for the purpose of
determining compliance with this part, it shall appear for purposes of section
801(a) of the act, that the methods used in, and the facilities and controls
used for, the design, manufacture, packaging, labeling, storage, installation,
or servicing of any devices produced at such facility that are offered for
import into the United States do not conform to the requirements of section
520(f) of the act and this part and that the devices manufactured at that
facility are adulterated under section 501(h) of the act. (e)Exemptions
or variances . (1) Any person who wishes to petition for an exemption or
variance from any device quality system requirement is subject to the
requirements of section 520(f)(2) of the act. Petitions for an exemption or
variance shall be submitted according to the procedures set forth in 10.30 of
this chapter, the FDA's administrative procedures. Guidance is available from
the Center for Devices and Radiological Health, Division of Small
Manufacturers, International and Consumer Assistance (HFZ-220), 1350 Piccard
Dr., Rockville, MD 20850, U.S.A., telephone 1-800-638-2041 or 240-276-3150, FAX
240-276-3151. (2) FDA
may initiate and grant a variance from any device quality system requirement
when the agency determines that such variance is in the best interest of the
public health. Such variance will remain in effect only so long as there
remains a public health need for the device and the device would not likely be
made sufficiently available without the variance. [61 FR
52654, Oct. 7, 1996, as amended at 65 FR 17136, Mar. 31, 2000; 65 FR 66636,
Nov. 7, 2000; 69 FR 29829, May 25, 2005; 72 FR 17399, Apr. 9, 2007] Sec.
820.3 Definitions. (a)Act
means the Federal Food, Drug, and Cosmetic Act, as amended (secs. 201-903, 52
Stat. 1040et seq., as amended (21 U.S.C. 321-394)). All definitions in section
201 of the act shall apply to the regulations in this part. (b)Complaint
means any written, electronic, or oral communication that alleges deficiencies
related to the identity, quality, durability, reliability, safety,
effectiveness, or performance of a device after it is released for
distribution. (c)Component
means any raw material, substance, piece, part, software, firmware, labeling,
or assembly which is intended to be included as part of the finished, packaged,
and labeled device. (d)Control
number means any distinctive symbols, such as a distinctive combination of
letters or numbers, or both, from which the history of the manufacturing,
packaging, labeling, and distribution of a unit, lot, or batch of finished
devices can be determined. (e)Design
history file (DHF ) means a compilation of records which describes the design
history of a finished device. (f)Design
input means the physical and performance requirements of a device that are used
as a basis for device design. (g)Design
output means the results of a design effort at each design phase and at the end
of the total design effort. The finished design output is the basis for the
device master record. The total finished design output consists of the device,
its packaging and labeling, and the device master record. (h)Design
review means a documented, comprehensive, systematic examination of a design to
evaluate the adequacy of the design requirements, to evaluate the capability of
the design to meet these requirements, and to identify problems. (i)Device
history record (DHR ) means a compilation of records containing the production
history of a finished device. (j)Device
master record (DMR ) means a compilation of records containing the procedures
and specifications for a finished device. (k)Establish
means define, document (in writing or electronically), and implement. (l)Finished
device means any device or accessory to any device that is suitable for use or
capable of functioning, whether or not it is packaged, labeled, or sterilized. (m)Lot or
batch means one or more components or finished devices that consist of a single
type, model, class, size, composition, or software version that are
manufactured under essentially the same conditions and that are intended to
have uniform characteristics and quality within specified limits. (n)Management
with executive responsibility means those senior employees of a manufacturer
who have the authority to establish or make changes to the manufacturer's
quality policy and quality system. (o)Manufacturer
means any person who designs, manufactures, fabricates, assembles, or processes
a finished device. Manufacturer includes but is not limited to those who
perform the functions of contract sterilization, installation, relabeling,
remanufacturing, repacking, or specification development, and initial
distributors of foreign entities performing these functions. (p)Manufacturing
material means any material or substance used in or used to facilitate the
manufacturing process, a concomitant constituent, or a byproduct constituent
produced during the manufacturing process, which is present in or on the
finished device as a residue or impurity not by design or intent of the
manufacturer. (q)Nonconformity
means the nonfulfillment of a specified requirement. (r)Product
means components, manufacturing materials, in- process devices, finished devices,
and returned devices. (s)Quality
means the totality of features and characteristics that bear on the ability of
a device to satisfy fitness-for-use, including safety and performance. (t)Quality
audit means a systematic, independent examination of a manufacturer's quality
system that is performed at defined intervals and at sufficient frequency to
determine whether both quality system activities and the results of such
activities comply with quality system procedures, that these procedures are
implemented effectively, and that these procedures are suitable to achieve
quality system objectives. (u)Quality
policy means the overall intentions and direction of an organization with
respect to quality, as established by management with executive responsibility. (v)Quality
system means the organizational structure, responsibilities, procedures,
processes, and resources for implementing quality management. (w)Remanufacturer
means any person who processes, conditions, renovates, repackages, restores, or
does any other act to a finished device that significantly changes the finished
device's performance or safety specifications, or intended use. (x)Rework
means action taken on a nonconforming product so that it will fulfill the
specified DMR requirements before it is released for distribution. (y)Specification
means any requirement with which a product, process, service, or other activity
must conform. (z)Validation
means confirmation by examination and provision of objective evidence that the
particular requirements for a specific intended use can be consistently
fulfilled. (1)Process
validation means establishing by objective evidence that a process consistently
produces a result or product meeting its predetermined specifications. (2)Design
validation means establishing by objective evidence that device specifications
conform with user needs and intended use(s). (aa)Verification
means confirmation by examination and provision of objective evidence that
specified requirements have been fulfilled. Sec.
820.5 Quality system. Each
manufacturer shall establish and maintain a quality system that is appropriate
for the specific medical device(s) designed or manufactured, and that meets the
requirements of this part. Subpart
B--Quality System Requirements Sec.
820.20 Management responsibility. (a)Quality
policy. Management with executive responsibility shall establish its policy and
objectives for, and commitment to, quality. Management with executive
responsibility shall ensure that the quality policy is understood, implemented,
and maintained at all levels of the organization. (b)Organization.
Each manufacturer shall establish and maintain an adequate organizational
structure to ensure that devices are designed and produced in accordance with
the requirements of this part. (1)Responsibility
and authority. Each manufacturer shall establish the appropriate
responsibility, authority, and interrelation of all personnel who manage,
perform, and assess work affecting quality, and provide the independence and
authority necessary to perform these tasks. (2)Resources.
Each manufacturer shall provide adequate resources, including the assignment of
trained personnel, for management, performance of work, and assessment
activities, including internal quality audits, to meet the requirements of this
part. (3)Management
representative. Management with executive responsibility shall appoint, and
document such appointment of, a member of management who, irrespective of other
responsibilities, shall have established authority over and responsibility for: (i)
Ensuring that quality system requirements are effectively established and
effectively maintained in accordance with this part; and (ii)
Reporting on the performance of the quality system to management with executive
responsibility for review. (c)Management
review. Management with executive responsibility shall review the suitability
and effectiveness of the quality system at defined intervals and with
sufficient frequency according to established procedures to ensure that the
quality system satisfies the requirements of this part and the manufacturer's
established quality policy and objectives. The dates and results of quality
system reviews shall be documented. (d)Quality
planning. Each manufacturer shall establish a quality plan which defines the
quality practices, resources, and activities relevant to devices that are
designed and manufactured. The manufacturer shall establish how the
requirements for quality will be met. (e)Quality
system procedures. Each manufacturer shall establish quality system procedures
and instructions. An outline of the structure of the documentation used in the
quality system shall be established where appropriate. Sec.
820.22 Quality audit. Each
manufacturer shall establish procedures for quality audits and conduct such
audits to assure that the quality system is in compliance with the established
quality system requirements and to determine the effectiveness of the quality
system. Quality audits shall be conducted by individuals who do not have direct
responsibility for the matters being audited. Corrective action(s), including a
reaudit of deficient matters, shall be taken when necessary. A report of the
results of each quality audit, and reaudit(s) where taken, shall be made and
such reports shall be reviewed by management having responsibility for the
matters audited. The dates and results of quality audits and reaudits shall be
documented. Sec.
820.25 Personnel. (a)General.
Each manufacturer shall have sufficient personnel with the necessary education,
background, training, and experience to assure that all activities required by
this part are correctly performed. (b)Training.
Each manufacturer shall establish procedures for identifying training needs and
ensure that all personnel are trained to adequately perform their assigned
responsibilities. Training shall be documented. (1) As
part of their training, personnel shall be made aware of device defects which
may occur from the improper performance of their specific jobs. (2)
Personnel who perform verification and validation activities shall be made
aware of defects and errors that may be encountered as part of their job
functions. Subpart
C--Design Controls Sec.
820.30 Design controls. (a)General.
(1) Each manufacturer of any class III or class II device, and the class I
devices listed in paragraph (a)(2) of this section, shall establish and
maintain procedures to control the design of the device in order to ensure that
specified design requirements are met. (2) The
following class I devices are subject to design controls: (i)
Devices automated with computer software; and (ii) The
devices listed in the following chart. Section Device 868.6810 Catheter, Tracheobronchial Suction. 878.4460 Glove, Surgeon's. 880.6760 Restraint, Protective. 892.5650 System, Applicator, Radionuclide,
Manual. 892.5740 Source, Radionuclide Teletherapy. (b)Design
and development planning. Each manufacturer shall establish and maintain plans
that describe or reference the design and development activities and define
responsibility for implementation. The plans shall identify and describe the
interfaces with different groups or activities that provide, or result in,
input to the design and development process. The plans shall be reviewed,
updated, and approved as design and development evolves. (c)Design
input. Each manufacturer shall establish and maintain procedures to ensure that
the design requirements relating to a device are appropriate and address the
intended use of the device, including the needs of the user and patient. The
procedures shall include a mechanism for addressing incomplete, ambiguous, or
conflicting requirements. The design input requirements shall be documented and
shall be reviewed and approved by a designated individual(s). The approval,
including the date and signature of the individual(s) approving the
requirements, shall be documented. (d)Design
output. Each manufacturer shall establish and maintain procedures for defining
and documenting design output in terms that allow an adequate evaluation of
conformance to design input requirements. Design output procedures shall
contain or make reference to acceptance criteria and shall ensure that those
design outputs that are essential for the proper functioning of the device are
identified. Design output shall be documented, reviewed, and approved before
release. The approval, including the date and signature of the individual(s)
approving the output, shall be documented. (e)Design
review. Each manufacturer shall establish and maintain procedures to ensure
that formal documented reviews of the design results are planned and conducted
at appropriate stages of the device's design development. The procedures shall
ensure that participants at each design review include representatives of all
functions concerned with the design stage being reviewed and an individual(s)
who does not have direct responsibility for the design stage being reviewed, as
well as any specialists needed. The results of a design review, including
identification of the design, the date, and the individual(s) performing the
review, shall be documented in the design history file (the DHF). (f)Design
verification. Each manufacturer shall establish and maintain procedures for
verifying the device design. Design verification shall confirm that the design
output meets the design input requirements. The results of the design verification,
including identification of the design, method(s), the date, and the
individual(s) performing the verification, shall be documented in the DHF. (g)Design
validation. Each manufacturer shall establish and maintain procedures for
validating the device design. Design validation shall be performed under
defined operating conditions on initial production units, lots, or batches, or
their equivalents. Design validation shall ensure that devices conform to
defined user needs and intended uses and shall include testing of production
units under actual or simulated use conditions. Design validation shall include
software validation and risk analysis, where appropriate. The results of the
design validation, including identification of the design, method(s), the date,
and the individual(s) performing the validation, shall be documented in the
DHF. (h)Design
transfer. Each manufacturer shall establish and maintain procedures to ensure
that the device design is correctly translated into production specifications. (i)Design
changes. Each manufacturer shall establish and maintain procedures for the
identification, documentation, validation or where appropriate verification,
review, and approval of design changes before their implementation. (j)Design
history file. Each manufacturer shall establish and maintain a DHF for each
type of device. The DHF shall contain or reference the records necessary to
demonstrate that the design was developed in accordance with the approved
design plan and the requirements of this part. Subpart
D--Document Controls Sec.
820.40 Document controls. Each
manufacturer shall establish and maintain procedures to control all documents
that are required by this part. The procedures shall provide for the following: (a)Document
approval and distribution. Each manufacturer shall designate an individual(s)
to review for adequacy and approve prior to issuance all documents established
to meet the requirements of this part. The approval, including the date and
signature of the individual(s) approving the document, shall be documented.
Documents established to meet the requirements of this part shall be available
at all locations for which they are designated, used, or otherwise necessary,
and all obsolete documents shall be promptly removed from all points of use or
otherwise prevented from unintended use. (b)Document
changes. Changes to documents shall be reviewed and approved by an
individual(s) in the same function or organization that performed the original
review and approval, unless specifically designated otherwise. Approved changes
shall be communicated to the appropriate personnel in a timely manner. Each
manufacturer shall maintain records of changes to documents. Change records
shall include a description of the change, identification of the affected
documents, the signature of the approving individual(s), the approval date, and
when the change becomes effective. Subpart
E--Purchasing Controls Sec.
820.50 Purchasing controls. Each
manufacturer shall establish and maintain procedures to ensure that all
purchased or otherwise received product and services conform to specified
requirements. (a)Evaluation
of suppliers, contractors, and consultants. Each manufacturer shall establish
and maintain the requirements, including quality requirements, that must be met
by suppliers, contractors, and consultants. Each manufacturer shall: (1)
Evaluate and select potential suppliers, contractors, and consultants on the
basis of their ability to meet specified requirements, including quality
requirements. The evaluation shall be documented. (2)
Define the type and extent of control to be exercised over the product,
services, suppliers, contractors, and consultants, based on the evaluation
results. (3)
Establish and maintain records of acceptable suppliers, contractors, and
consultants. (b)Purchasing
data. Each manufacturer shall establish and maintain data that clearly describe
or reference the specified requirements, including quality requirements, for
purchased or otherwise received product and services. Purchasing documents
shall include, where possible, an agreement that the suppliers, contractors,
and consultants agree to notify the manufacturer of changes in the product or
service so that manufacturers may determine whether the changes may affect the
quality of a finished device. Purchasing data shall be approved in accordance
with 820.40. Subpart
F--Identification and Traceability Sec.
820.60 Identification. Each
manufacturer shall establish and maintain procedures for identifying product
during all stages of receipt, production, distribution, and installation to
prevent mixups. Sec.
820.65 Traceability. Each
manufacturer of a device that is intended for surgical implant into the body or
to support or sustain life and whose failure to perform when properly used in
accordance with instructions for use provided in the labeling can be reasonably
expected to result in a significant injury to the user shall establish and
maintain procedures for identifying with a control number each unit, lot, or
batch of finished devices and where appropriate components. The procedures
shall facilitate corrective action. Such identification shall be documented in
the DHR. Subpart
G--Production and Process Controls Sec.
820.70 Production and process controls. (a)General.
Each manufacturer shall develop, conduct, control, and monitor production
processes to ensure that a device conforms to its specifications. Where
deviations from device specifications could occur as a result of the
manufacturing process, the manufacturer shall establish and maintain process
control procedures that describe any process controls necessary to ensure
conformance to specifications. Where process controls are needed they shall
include: (1)
Documented instructions, standard operating procedures (SOP's), and methods
that define and control the manner of production; (2)
Monitoring and control of process parameters and component and device
characteristics during production; (3)
Compliance with specified reference standards or codes; (4) The
approval of processes and process equipment; and (5)
Criteria for workmanship which shall be expressed in documented standards or by
means of identified and approved representative samples. (b)Production
and process changes. Each manufacturer shall establish and maintain procedures
for changes to a specification, method, process, or procedure. Such changes
shall be verified or where appropriate validated according to 820.75, before
implementation and these activities shall be documented. Changes shall be
approved in accordance with 820.40. (c)Environmental
control. Where environmental conditions could reasonably be expected to have an
adverse effect on product quality, the manufacturer shall establish and
maintain procedures to adequately control these environmental conditions.
Environmental control system(s) shall be periodically inspected to verify that
the system, including necessary equipment, is adequate and functioning
properly. These activities shall be documented and reviewed. (d)Personnel.
Each manufacturer shall establish and maintain requirements for the health,
cleanliness, personal practices, and clothing of personnel if contact between
such personnel and product or environment could reasonably be expected to have
an adverse effect on product quality. The manufacturer shall ensure that
maintenance and other personnel who are required to work temporarily under
special environmental conditions are appropriately trained or supervised by a
trained individual. (e)Contamination
control. Each manufacturer shall establish and maintain procedures to prevent
contamination of equipment or product by substances that could reasonably be
expected to have an adverse effect on product quality. (f)Buildings.
Buildings shall be of suitable design and contain sufficient space to perform
necessary operations, prevent mixups, and assure orderly handling. (g)Equipment.
Each manufacturer shall ensure that all equipment used in the manufacturing
process meets specified requirements and is appropriately designed,
constructed, placed, and installed to facilitate maintenance, adjustment,
cleaning, and use. (1)Maintenance
schedule. Each manufacturer shall establish and maintain schedules for the
adjustment, cleaning, and other maintenance of equipment to ensure that
manufacturing specifications are met. Maintenance activities, including the
date and individual(s) performing the maintenance activities, shall be
documented. (2)Inspection.
Each manufacturer shall conduct periodic inspections in accordance with established
procedures to ensure adherence to applicable equipment maintenance schedules.
The inspections, including the date and individual(s) conducting the
inspections, shall be documented. (3)Adjustment.
Each manufacturer shall ensure that any inherent limitations or allowable
tolerances are visibly posted on or near equipment requiring periodic
adjustments or are readily available to personnel performing these adjustments. (h)Manufacturing
material. Where a manufacturing material could reasonably be expected to have
an adverse effect on product quality, the manufacturer shall establish and
maintain procedures for the use and removal of such manufacturing material to
ensure that it is removed or limited to an amount that does not adversely
affect the device's quality. The removal or reduction of such manufacturing
material shall be documented. (i)Automated
processes. When computers or automated data processing systems are used as part
of production or the quality system, the manufacturer shall validate computer
software for its intended use according to an established protocol. All
software changes shall be validated before approval and issuance. These
validation activities and results shall be documented. Sec.
820.72 Inspection, measuring, and test equipment. (a)Control
of inspection, measuring, and test equipment. Each manufacturer shall ensure
that all inspection, measuring, and test equipment, including mechanical,
automated, or electronic inspection and test equipment, is suitable for its
intended purposes and is capable of producing valid results. Each manufacturer
shall establish and maintain procedures to ensure that equipment is routinely
calibrated, inspected, checked, and maintained. The procedures shall include
provisions for handling, preservation, and storage of equipment, so that its
accuracy and fitness for use are maintained. These activities shall be
documented. (b)Calibration.
Calibration procedures shall include specific directions and limits for
accuracy and precision. When accuracy and precision limits are not met, there
shall be provisions for remedial action to reestablish the limits and to
evaluate whether there was any adverse effect on the device's quality. These
activities shall be documented. (1)Calibration
standards. Calibration standards used for inspection, measuring, and test
equipment shall be traceable to national or international standards. If
national or international standards are not practical or available, the
manufacturer shall use an independent reproducible standard. If no applicable
standard exists, the manufacturer shall establish and maintain an in-house
standard. (2)Calibration
records. The equipment identification, calibration dates, the individual
performing each calibration, and the next calibration date shall be documented.
These records shall be displayed on or near each piece of equipment or shall be
readily available to the personnel using such equipment and to the individuals
responsible for calibrating the equipment. Sec.
820.75 Process validation. (a) Where
the results of a process cannot be fully verified by subsequent inspection and
test, the process shall be validated with a high degree of assurance and
approved according to established procedures. The validation activities and
results, including the date and signature of the individual(s) approving the
validation and where appropriate the major equipment validated, shall be
documented. (b) Each
manufacturer shall establish and maintain procedures for monitoring and control
of process parameters for validated processes to ensure that the specified
requirements continue to be met. (1) Each
manufacturer shall ensure that validated processes are performed by qualified
individual(s). (2) For
validated processes, the monitoring and control methods and data, the date
performed, and, where appropriate, the individual(s) performing the process or
the major equipment used shall be documented. (c) When
changes or process deviations occur, the manufacturer shall review and evaluate
the process and perform revalidation where appropriate. These activities shall
be documented. Subpart
H--Acceptance Activities Sec.
820.80 Receiving, in-process, and finished device acceptance. (a)General.
Each manufacturer shall establish and maintain procedures for acceptance
activities. Acceptance activities include inspections, tests, or other
verification activities. (b)Receiving
acceptance activities. Each manufacturer shall establish and maintain
procedures for acceptance of incoming product. Incoming product shall be
inspected, tested, or otherwise verified as conforming to specified
requirements. Acceptance or rejection shall be documented. (c)In-process
acceptance activities. Each manufacturer shall establish and maintain
acceptance procedures, where appropriate, to ensure that specified requirements
for in-process product are met. Such procedures shall ensure that in-process
product is controlled until the required inspection and tests or other
verification activities have been completed, or necessary approvals are
received, and are documented. (d)Final
acceptance activities. Each manufacturer shall establish and maintain
procedures for finished device acceptance to ensure that each production run,
lot, or batch of finished devices meets acceptance criteria. Finished devices
shall be held in quarantine or otherwise adequately controlled until released.
Finished devices shall not be released for distribution until: (1) The
activities required in the DMR are completed; (2) the
associated data and documentation is reviewed; (3) the
release is authorized by the signature of a designated individual(s); and (4) the authorization
is dated. (e)Acceptance
records. Each manufacturer shall document acceptance activities required by
this part. These records shall include: (1) The
acceptance activities performed; (2) the
dates acceptance activities are performed; (3) the
results; (4) the
signature of the individual(s) conducting the acceptance activities; and (5) where
appropriate the equipment used. These records shall be part of the DHR. Sec.
820.86 Acceptance status. Each
manufacturer shall identify by suitable means the acceptance status of product,
to indicate the conformance or nonconformance of product with acceptance
criteria. The identification of acceptance status shall be maintained
throughout manufacturing, packaging, labeling, installation, and servicing of
the product to ensure that only product which has passed the required
acceptance activities is distributed, used, or installed. Subpart
I--Nonconforming Product Sec.
820.90 Nonconforming product. (a)Control
of nonconforming product. Each manufacturer shall establish and maintain
procedures to control product that does not conform to specified requirements.
The procedures shall address the identification, documentation, evaluation,
segregation, and disposition of nonconforming product. The evaluation of
nonconformance shall include a determination of the need for an investigation
and notification of the persons or organizations responsible for the
nonconformance. The evaluation and any investigation shall be documented. (b)Nonconformity
review and disposition. (1) Each manufacturer shall establish and maintain
procedures that define the responsibility for review and the authority for the
disposition of nonconforming product. The procedures shall set forth the review
and disposition process. Disposition of nonconforming product shall be
documented. Documentation shall include the justification for use of
nonconforming product and the signature of the individual(s) authorizing the
use. (2) Each
manufacturer shall establish and maintain procedures for rework, to include
retesting and reevaluation of the nonconforming product after rework, to ensure
that the product meets its current approved specifications. Rework and
reevaluation activities, including a determination of any adverse effect from
the rework upon the product, shall be documented in the DHR. Subpart
J--Corrective and Preventive Action Sec.
820.100 Corrective and preventive action. (a) Each
manufacturer shall establish and maintain procedures for implementing
corrective and preventive action. The procedures shall include requirements
for: (1)
Analyzing processes, work operations, concessions, quality audit reports,
quality records, service records, complaints, returned product, and other sources
of quality data to identify existing and potential causes of nonconforming
product, or other quality problems. Appropriate statistical methodology shall
be employed where necessary to detect recurring quality problems; (2)
Investigating the cause of nonconformities relating to product, processes, and
the quality system; (3)
Identifying the action(s) needed to correct and prevent recurrence of
nonconforming product and other quality problems; (4)
Verifying or validating the corrective and preventive action to ensure that
such action is effective and does not adversely affect the finished device; (5)
Implementing and recording changes in methods and procedures needed to correct
and prevent identified quality problems; (6)
Ensuring that information related to quality problems or nonconforming product
is disseminated to those directly responsible for assuring the quality of such
product or the prevention of such problems; and (7)
Submitting relevant information on identified quality problems, as well as
corrective and preventive actions, for management review. (b) All
activities required under this section, and their results, shall be documented. Subpart
K--Labeling and Packaging Control Sec.
820.120 Device labeling. Each
manufacturer shall establish and maintain procedures to control labeling
activities. (a)Label
integrity. Labels shall be printed and applied so as to remain legible and
affixed during the customary conditions of processing, storage, handling,
distribution, and where appropriate use. (b)Labeling
inspection. Labeling shall not be released for storage or use until a
designated individual(s) has examined the labeling for accuracy including,
where applicable, the correct expiration date, control number, storage
instructions, handling instructions, and any additional processing
instructions. The release, including the date and signature of the
individual(s) performing the examination, shall be documented in the DHR. (c)Labeling
storage. Each manufacturer shall store labeling in a manner that provides
proper identification and is designed to prevent mixups. (d)Labeling
operations. Each manufacturer shall control labeling and packaging operations
to prevent labeling mixups. The label and labeling used for each production
unit, lot, or batch shall be documented in the DHR. (e)Control
number. Where a control number is required by 820.65, that control number shall
be on or shall accompany the device through distribution. Sec.
820.130 Device packaging. Each manufacturer
shall ensure that device packaging and shipping containers are designed and
constructed to protect the device from alteration or damage during the
customary conditions of processing, storage, handling, and distribution. Subpart
L--Handling, Storage, Distribution, and Installation Sec.
820.140 Handling. Each
manufacturer shall establish and maintain procedures to ensure that mixups,
damage, deterioration, contamination, or other adverse effects to product do
not occur during handling. Sec.
820.150 Storage. (a) Each
manufacturer shall establish and maintain procedures for the control of storage
areas and stock rooms for product to prevent mixups, damage, deterioration,
contamination, or other adverse effects pending use or distribution and to ensure
that no obsolete, rejected, or deteriorated product is used or distributed.
When the quality of product deteriorates over time, it shall be stored in a
manner to facilitate proper stock rotation, and its condition shall be assessed
as appropriate. (b) Each
manufacturer shall establish and maintain procedures that describe the methods
for authorizing receipt from and dispatch to storage areas and stock rooms. Sec.
820.160 Distribution. (a) Each
manufacturer shall establish and maintain procedures for control and
distribution of finished devices to ensure that only those devices approved for
release are distributed and that purchase orders are reviewed to ensure that
ambiguities and errors are resolved before devices are released for
distribution. Where a device's fitness for use or quality deteriorates over
time, the procedures shall ensure that expired devices or devices deteriorated
beyond acceptable fitness for use are not distributed. (b) Each
manufacturer shall maintain distribution records which include or refer to the
location of: (1) The
name and address of the initial consignee; (2) The
identification and quantity of devices shipped; (3) The
date shipped; and (4) Any
control number(s) used. Sec.
820.170 Installation. (a) Each
manufacturer of a device requiring installation shall establish and maintain
adequate installation and inspection instructions, and where appropriate test
procedures. Instructions and procedures shall include directions for ensuring
proper installation so that the device will perform as intended after
installation. The manufacturer shall distribute the instructions and procedures
with the device or otherwise make them available to the person(s) installing
the device. (b) The
person installing the device shall ensure that the installation, inspection,
and any required testing are performed in accordance with the manufacturer's
instructions and procedures and shall document the inspection and any test
results to demonstrate proper installation. Subpart
M--Records Sec.
820.180 General requirements. All
records required by this part shall be maintained at the manufacturing
establishment or other location that is reasonably accessible to responsible
officials of the manufacturer and to employees of FDA designated to perform
inspections. Such records, including those not stored at the inspected
establishment, shall be made readily available for review and copying by FDA
employee(s). Such records shall be legible and shall be stored to minimize
deterioration and to prevent loss. Those records stored in automated data
processing systems shall be backed up. (a)Confidentiality.
Records deemed confidential by the manufacturer may be marked to aid FDA in
determining whether information may be disclosed under the public information
regulation in part 20 of this chapter. (b)Record
retention period. All records required by this part shall be retained for a
period of time equivalent to the design and expected life of the device, but in
no case less than 2 years from the date of release for commercial distribution
by the manufacturer. (c)Exceptions.
This section does not apply to the reports required by 820.20(c) Management
review, 820.22 Quality audits, and supplier audit reports used to meet the
requirements of 820.50(a) Evaluation of suppliers, contractors, and
consultants, but does apply to procedures established under these provisions.
Upon request of a designated employee of FDA, an employee in management with
executive responsibility shall certify in writing that the management reviews
and quality audits required under this part, and supplier audits where
applicable, have been performed and documented, the dates on which they were
performed, and that any required corrective action has been undertaken. Sec.
820.181 Device master record. Each
manufacturer shall maintain device master records (DMR's). Each manufacturer
shall ensure that each DMR is prepared and approved in accordance with 820.40.
The DMR for each type of device shall include, or refer to the location of, the
following information: (a)
Device specifications including appropriate drawings, composition, formulation,
component specifications, and software specifications; (b)
Production process specifications including the appropriate equipment
specifications, production methods, production procedures, and production
environment specifications; (c)
Quality assurance procedures and specifications including acceptance criteria
and the quality assurance equipment to be used; (d)
Packaging and labeling specifications, including methods and processes used;
and (e)
Installation, maintenance, and servicing procedures and methods. Sec.
820.184 Device history record. Each
manufacturer shall maintain device history records (DHR's). Each manufacturer
shall establish and maintain procedures to ensure that DHR's for each batch,
lot, or unit are maintained to demonstrate that the device is manufactured in
accordance with the DMR and the requirements of this part. The DHR shall
include, or refer to the location of, the following information: (a) The
dates of manufacture; (b) The
quantity manufactured; (c) The
quantity released for distribution; (d) The
acceptance records which demonstrate the device is manufactured in accordance
with the DMR; (e) The
primary identification label and labeling used for each production unit; and (f) Any
device identification(s) and control number(s) used. Sec.
820.186 Quality system record. Each
manufacturer shall maintain a quality system record (QSR). The QSR shall
include, or refer to the location of, procedures and the documentation of
activities required by this part that are not specific to a particular type of
device(s), including, but not limited to, the records required by 820.20. Each
manufacturer shall ensure that the QSR is prepared and approved in accordance
with 820.40. Sec.
820.198 Complaint files. (a) Each
manufacturer shall maintain complaint files. Each manufacturer shall establish
and maintain procedures for receiving, reviewing, and evaluating complaints by
a formally designated unit. Such procedures shall ensure that: (1) All
complaints are processed in a uniform and timely manner; (2) Oral
complaints are documented upon receipt; and (3)
Complaints are evaluated to determine whether the complaint represents an event
which is required to be reported to FDA under part 803 of this chapter, Medical
Device Reporting. (b) Each
manufacturer shall review and evaluate all complaints to determine whether an
investigation is necessary. When no investigation is made, the manufacturer
shall maintain a record that includes the reason no investigation was made and
the name of the individual responsible for the decision not to investigate. (c) Any
complaint involving the possible failure of a device, labeling, or packaging to
meet any of its specifications shall be reviewed, evaluated, and investigated,
unless such investigation has already been performed for a similar complaint
and another investigation is not necessary. (d) Any
complaint that represents an event which must be reported to FDA under part 803
of this chapter shall be promptly reviewed, evaluated, and investigated by a
designated individual(s) and shall be maintained in a separate portion of the
complaint files or otherwise clearly identified. In addition to the information
required by 820.198(e), records of investigation under this paragraph shall
include a determination of: (1)
Whether the device failed to meet specifications; (2)
Whether the device was being used for treatment or diagnosis; and (3) The
relationship, if any, of the device to the reported incident or adverse event. (e) When
an investigation is made under this section, a record of the investigation shall
be maintained by the formally designated unit identified in paragraph (a) of
this section. The record of investigation shall include: (1) The
name of the device; (2) The
date the complaint was received; (3) Any
device identification(s) and control number(s) used; (4) The
name, address, and phone number of the complainant; (5) The
nature and details of the complaint; (6) The
dates and results of the investigation; (7) Any
corrective action taken; and (8) Any
reply to the complainant. (f) When
the manufacturer's formally designated complaint unit is located at a site
separate from the manufacturing establishment, the investigated complaint(s)
and the record(s) of investigation shall be reasonably accessible to the
manufacturing establishment. (g) If a
manufacturer's formally designated complaint unit is located outside of the
United States, records required by this section shall be reasonably accessible
in the United States at either: (1) A
location in the United States where the manufacturer's records are regularly
kept; or (2) The
location of the initial distributor. [61 FR
52654, Oct. 7, 1996, as amended at 69 FR 11313, Mar. 10, 2004; 71 FR 16228,
Mar. 31, 2006] Subpart
N--Servicing Sec.
820.200 Servicing. (a) Where
servicing is a specified requirement, each manufacturer shall establish and
maintain instructions and procedures for performing and verifying that the
servicing meets the specified requirements. (b) Each
manufacturer shall analyze service reports with appropriate statistical
methodology in accordance with 820.100. (c) Each
manufacturer who receives a service report that represents an event which must
be reported to FDA under part 803 of this chapter shall automatically consider
the report a complaint and shall process it in accordance with the requirements
of 820.198. (d)
Service reports shall be documented and shall include: (1) The
name of the device serviced; (2) Any
device identification(s) and control number(s) used; (3) The
date of service; (4) The
individual(s) servicing the device; (5) The
service performed; and (6) The
test and inspection data. [61 FR
52654, Oct. 7, 1996, as amended at 69 FR 11313, Mar. 10, 2004] Subpart
O--Statistical Techniques Sec.
820.250 Statistical techniques. (a) Where
appropriate, each manufacturer shall establish and maintain procedures for
identifying valid statistical techniques required for establishing,
controlling, and verifying the acceptability of process capability and product
characteristics. (b)
Sampling plans, when used, shall be written and based on a valid statistical
rationale. Each manufacturer shall establish and maintain procedures to ensure
that sampling methods are adequate for their intended use and to ensure that
when changes occur the sampling plans are reviewed. These activities shall be
documented. |
|